【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc

上传人:sccc 文档编号:5208155 上传时间:2023-06-13 格式:DOC 页数:13 大小:480KB
返回 下载 相关 举报
【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc_第1页
第1页 / 共13页
【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc_第2页
第2页 / 共13页
【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc_第3页
第3页 / 共13页
【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc_第4页
第4页 / 共13页
【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc_第5页
第5页 / 共13页
点击查看更多>>
资源描述

《【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc》由会员分享,可在线阅读,更多相关《【BS英国标准】BS EN 86831999 Packaging materials and systems for medical devices which are to be.doc(13页珍藏版)》请在三一办公上搜索。

1、BRITISH STANDARDPackaging materials and systems for medical devices which are to be sterilized Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) Requirements and test methodsThe European Standard EN 868-

2、3:1999 has the status of aBritish StandardICS 11.080.30; 55.040;| BS EN868-3:1999|Licensed Copy: London South Bank University, London South Bank University, Wed Jul 12 21:50:40 BST 2006, Uncontrolled Copy, (c) BSI|NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAWBS EN 868-3:1999

3、National forewordThis British Standard is the English language version of EN 868-3:1999. It supersedes most of the normative text of BS 6256:1989, which is subject to amendment to this effect. BS 6256:1989 is not yet withdrawn, because reference is made to it inannex D of EN 868-1:1997, with referen

4、ce to the possibility of applying the methylene blue particulate penetration test (described in appendix C of BS 6256) as a means of assessing the microbial barrier properties of some types of sterilization packaging covered by EN 868.The UK participation in its preparation was entrusted by Technica

5、l Committee LBI/35, Sterilizers, autoclaves and disinfectors, to Subcommittee LBI/35/9, Sterilization packaging materials, which has the responsibility to: aid enquirers to understand the text; present to the responsible European committee any enquiries on the interpretation, or proposals for change

6、, and keep the UK interests informed; monitor related international and European developments and promulgate them in the UK.A list of organizations represented on this subcommittee can be obtained on request to its secretary.Cross-referencesThe British Standards which implement international or Euro

7、pean publications referred to in this document may be found in the BSI Standards Catalogue under the section entitled International Standards Correspondence Index, or by using theFind facility of the BSI Standards Electronic Catalogue.Sterilization packaging materials falling within the scope of EN

8、868-1 are considered to be medical devices under Directive 93/42/EEC and compliance with the requirements of EN 868-1 is a means of ensuring that particular essential requirements of the Directive are met (see annex ZA of EN 868-1). This part ofEN 868 may be used to demonstrate compliance with some

9、of the requirements of Part 1. However, compliance with this part of EN 868 alone may not necessarily be considered to ensure that the essential requirements of the Directive have been met.A British Standard does not purport to include all the necessary provisions of a contract. Users of British Sta

10、ndards are responsible for their correct application.Compliance with a British Standard does not of itself confer immunity from legal obligations.Licensed Copy: London South Bank University, London South Bank University, Wed Jul 12 21:50:40 BST 2006, Uncontrolled Copy, (c) BSIThis British Standard,

11、having been prepared under the direction of the Sector Committee for Materials and Chemicals, was published under the authority of the StandardsCommittee and comes into effect on 15 November 1999BSI 11-1999ISBN 0 580 35316 8Summary of pagesThis document comprises a front cover, an inside front cover

12、, the EN title page, pages 2 to 8, an inside back cover and a back cover.The BSI copyright notice displayed in this document indicates when the document was last issued.Amd. No.DateCommentsAmendments issued since publicationEUROPEAN STANDARDEN 868-3NORME EUROPE ENNEEUROPA ISCHE NORMJune 1999ICS 11.0

13、80; 55.040English versionPackaging materials and systems for medical devices which are to be sterilized Part 3: Paper for use in themanufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) Requirements and test methodsMate riaux et syste m

14、es demballages pour les dispositifs me dicaux devant e tre ste rilise s Partie 3: Papier utilise dans la fabrication de sacs en papier (spe cifie s dans lEN 868-4) et dans la fabrication de sachets et gaines (spe cifie s danslEN 868-5) Exigences et me thodes dessaiVerpackungsmaterialien und -systeme

15、 fu r zu sterilisierende Medizinprodukte Teil 3: Papier zur Herstellung von Papierbeuteln (festgelegt inEN 868-4) und zur Herstellung von Klarsichtbeuteln und -schla uchen (festgelegt in EN 868-5) Anforderungen und Pru fverfahrenLicensed Copy: London South Bank University, London South Bank Universi

16、ty, Wed Jul 12 21:50:40 BST 2006, Uncontrolled Copy, (c) BSIThis European Standard was approved by CEN on 13 May 1999.CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without a

17、ny alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the Central Secretariat or to any CEN member.This European Standard exists in three official versions (English, French, German). A version in any other language made by

18、 translation under the responsibility of a CEN member into its own language and notified to the Central Secretariat has the same status as the official versions.CEN members are the national standards bodies of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, Irel

19、and, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and United Kingdom.CENEuropean Committee for Standardization Comite Europe en de Normalisation Europa isches Komitee fu r NormungCentral Secretariat: rue de Stassart 36, B-1050 Brussels1999 CEN All rights of exploitati

20、on in any form and by any means reserved worldwide for CEN nationalMembers.Ref. No. EN 868-3:1999 EPage 7EN 868-3:1999BSI 11-1999ForewordThis European Standard has been prepared byTechnical Committee CEN/TC 102, Sterilizers forContentsPageLicensed Copy: London South Bank University, London South Ban

21、k University, Wed Jul 12 21:50:40 BST 2006, Uncontrolled Copy, (c) BSImedical purposes, the Secretariat of which is held byDIN.This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest byDecember 1999, and con

22、flicting national standards shall be withdrawn at the latest by December 1999.This standard is one of a series of European Standards concerned with packaging materials and systems for medical devices which are to be sterilized. This series consists of the following parts.EN 868-1, Packaging material

23、s and systems for medical devices which are to be sterilized Part 1: General requirements and test methods. EN 868-2, Packaging materials and systems formedical devices which are to be sterilized Part 2: Sterilization wrap Requirements and test methods.EN 868-3, Packaging materials and systems for m

24、edical devices which are to be sterilized Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) Requirements and test methods.EN 868-4, Packaging materials and systems for medical devices which are to be ste

25、rilized Part 4: Paper bags Requirements and test methods.EN 868-5, Packaging materials and systems for medical devices which are to be sterilized Part 5: Heat and self sealable pouches and reels of paper and plastic film construction Requirements and test methods.EN 868-6, Packaging materials and sy

26、stems for medical devices which are to be sterilized Part 6: Paper for the manufacture of packs formedical use for sterilization by ethylene oxide or irradiation Requirements and test methods.EN 868-7, Packaging materials and systems for medical devices which are to be sterilized Part 7: Adhesive co

27、ated paper for the manufacture of heat sealable packs for medical use for sterilization by ethylene oxide or irradiation Requirements andtest methods.EN 868-8, Packaging materials and systems for medical devices which are to be sterilized Part 8: Re-usable sterilization containers for steam steriliz

28、ers conforming to EN 285 Requirements and test methods.According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Icelan

29、d, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and the United Kingdom.Foreword 2Introduction 31Scope 32Normative references 33Definitions 34Requirements 3Annex A (normative) Method for thedetermination of water repellency 5Annex B (normative) Method for thed

30、etermination of pore size 5Annex C (informative) Dimensions andtolerances 8Annex D (informative) Bibliography 8Licensed Copy: London South Bank University, London South Bank University, Wed Jul 12 21:50:40 BST 2006, Uncontrolled Copy, (c) BSIIntroductionPart 1 of this series of European Standards sp

31、ecifies general requirements and test methods for all packaging materials and systems intended for use as packaging for medical devices which are to be terminally sterilized in their packaging.1 ScopeThis part of the series of EN 868 provides examples of particular requirements and test methods for

32、paper, used in the manufacture of paper bags (specified inEN 868-4) and in the manufacture of pouches and reels(specified in EN 868-5).It introduces no additional requirements to the general requirements of Part 1 but provides guidance based upon various elements of former, relevant national standar

33、ds.As such, the particular requirements in 4.2 to 4.5 can be used to demonstrate compliance with one or more but not all of the requirements of Part 1.Paper specified in this part is suitable for use as packaging of medical devices which are to be terminally sterilized.2 Normative referencesThis Eur

34、opean Standard incorporates, by dated or undated reference, provisions from other publications. These normative references are cited at the appropriate places in the text and the publications are listed hereafter. For dated references, subsequent amendments to or revisions of any of these publicatio

35、ns apply to this European Standard onlywhen incorporated in it by amendment or revision. For undated references, the latest edition of the publication referred to applies.EN 868-1, Packaging materials and systems for medical devices which are to be sterilized Part 1: General requirements and test me

36、thods.EN 20187, Paper, board and pulps Standard atmosphere for conditioning and testing and procedure for monitoring the atmosphere and conditioning of samples.(ISO 187:1990)EN 20535, Paper and board Determination of water absorptiveness Cobb method.(ISO 535:1991)EN 21974, Paper Determination of tea

37、ring resistance (Elmendorf method).(ISO 1974:1990)EN 28601, Data elements and interchange formats Information interchange Representation of dates and times.(ISO 8601:1988 and technical corrigendum 1:1991)EN ISO 1924-2, Paper and board Determination of tensile properties Part 2: Constant rate of elon

38、gation method.(ISO 1924-2:1994)ISO 536, Paper and board Determination of grammage.ISO 2758, Paper Determination of bursting strength.ISO 3689, Paper and board Determination of bursting strength after immersion in water. (Revision of ISO 3689:1976)ISO 3781, Paper and board Determination of tensile st

39、rength after immersion in water.ISO 5636-3, Paper and board Determination of air permeance (medium range) Part 3: Bendtsen method.ISO 6588, Paper, board and pulps Determination of pH of aqueous extracts.ISO 9197-1, Paper, board and pulps Determination of water-soluble chlorides Part 1: General metho

40、d.ISO 9198, Paper, board and pulps Determination of water-soluble sulfates Titrimetric method.DIN 58953-6:1987, Sterilization Sterile supply Sterilization paper for bags and tube packings Test.3 DefinitionsFor the purposes of this European Standard, the definitions of EN 868-1 apply.4 Requirements4.

41、1 GeneralThe requirements of EN 868-1 apply.NOTE The following particular requirements and test methods can be used to demonstrate compliance with one or more, but not all, of the requirements of EN 868-1.4.2 Materials4.2.1 Raw materials used for the manufacture of packaging materials may be virgin

42、or reclaimed materials provided that the source, history and traceability of all raw materials, especially recycled materials, are known and controlled to ensure that the finished product will consistently meet the requirements of this standard.NOTE With current commercial technologies it is unlikel

43、y that reclaimed material other than manufacturing waste will be sufficiently controlled to allow its safe use for medical device packaging.4.2.2 The pulp shall be free from grit and untreated fragments of the original materials from which the pulp was prepared.4.3 ConditioningWhere reference is mad

44、e in the following performance requirements to conditioned paper or to the need for conditioning prior to carrying out a test the paper shall be conditioned in accordance with the method given in EN 20187.4.4 Performance requirements and test methodsNOTE The requirements of 4.4.12, 4.4.13 and 4.4.16

45、 do not apply to materials solely for use in irradiation sterilization packaging.4.4.1 When examined by unaided normal or corrected vision in transmitted light (daylight or good artificial light) the paper shall be free from tears, creases or localized thickening sufficient to impair its functioning.4.4.2 No colour shall leach out of the paper. Compli